Aiyaku

Dentitude ™ U-Zee

[Product Name]]Dental Lithium Disilicate Glass Ceramics

[Model Specification]]

Model: HT-0M1, HT-0M2, HT-0M3, HT-0M3,HT-A1, HT-A2, HT-A3, HT-B1, HT-B2

LT-0M1, LT-0M2, LT-0M3, LT-0M3, LT-A1, LT-A2, LT-A3, LT-B1, LT-B2

Specifications:L-18-15-13

[Type Classification]]This product belongs to Type II Class 2 dental ceramic materials, used to make veneers, inlays, onlays, single unit anterior teeth and posterior teeth restorations.

[Scope Of Application]]

The use of CAD/CAM technology to produce veneer, inlay, onlay, adhesive retention of single unit anterior and posterior tooth restorations.

[Method Of Use]]

Cutting: without sintering, directly cut into the required restoration body, glaze can be used.

 

Glazing Procedure


Preheating

2 Minutes.

Starting Temperature

500 ℃

Heating Rate

50℃/min

Final Temperature

800 ℃

Vacuum

Close

Hold Time

1 Minute

 

Products can be toned with dyeing solution. It is recommended to use commercially available products that have obtained medical device registration and production license and are suitable for dyeing lithium-based glass ceramics, such as IPS e.max Ceram Glaze Paste, VITA AKZENT Plus GLAZE LT and other series of dyes produced by Yiwujia.

Physical and Chemical Properties]

  • Glass-ceramics after sintering: 200±30 MPa;

  • Glass ceramics after sintering: 450 ℃ ± 20 ℃;

  • Glass-ceramics after sintering:(9.8±0.5)× 10-6 K-1;

  • Glass ceramics after sintering: ≤ 100 & micro;g/cm2;

  • Glass-ceramics after sintering: uranium -238 activity concentration ≤ 1.0 Bq/g.

Biological Performance]

In vitro cytotoxicity test: cytotoxicity is not greater than grade 1;

Delayed hypersensitivity test: no delayed hypersensitivity;

Intracutmal reaction test: the difference between the average score of the test sample and the solvent control is 0;

Acute systemic toxicity test: no acute systemic toxicity;

Sub-chronic systemic toxicity test: no sub-chronic systemic toxicity

Amestest: negative for mutation

Mouse lymphoma gene mutation test: mutagenesis was negative.

[Applicable equipment]]This product is suitable for dental cutting or die casting equipment.

Contraindications]None

Storage Conditions]were stored at room temperature.

[Validity]]10 years.

[Note]]

1.Is made by professional technicians and used by professional doctors, and the products are in contact with oral mucosa after processing;

2. Do not inhale ceramic dust during the trimming process, and timely discharge exhaust gas and wear protective measures such as masks.


PRE:Period NEXT:Dentitude™ Amber Porcelain