[Product Name]]Dental Lithium Disilicate Glass Ceramics
[Model Specification]]
Model: HT-0M1, HT-0M2, HT-0M3, HT-0M3,HT-A1, HT-A2, HT-A3, HT-B1, HT-B2
LT-0M1, LT-0M2, LT-0M3, LT-0M3, LT-A1, LT-A2, LT-A3, LT-B1, LT-B2
Specifications:L-18-15-13
[Type Classification]]This product belongs to Type II Class 2 dental ceramic materials, used to make veneers, inlays, onlays, single unit anterior teeth and posterior teeth restorations.
[Scope Of Application]]
The use of CAD/CAM technology to produce veneer, inlay, onlay, adhesive retention of single unit anterior and posterior tooth restorations.
[Method Of Use]]
Cutting: without sintering, directly cut into the required restoration body, glaze can be used.
Glazing Procedure | |
Preheating | 2 Minutes. |
Starting Temperature | 500 ℃ |
Heating Rate | 50℃/min |
Final Temperature | 800 ℃ |
Vacuum | Close |
Hold Time | 1 Minute |
Products can be toned with dyeing solution. It is recommended to use commercially available products that have obtained medical device registration and production license and are suitable for dyeing lithium-based glass ceramics, such as IPS e.max Ceram Glaze Paste, VITA AKZENT Plus GLAZE LT and other series of dyes produced by Yiwujia.
Physical and Chemical Properties]
Glass-ceramics after sintering: 200±30 MPa;
Glass ceramics after sintering: 450 ℃ ± 20 ℃;
Glass-ceramics after sintering:(9.8±0.5)× 10-6 K-1;
Glass ceramics after sintering: ≤ 100 & micro;g/cm2;
Glass-ceramics after sintering: uranium -238 activity concentration ≤ 1.0 Bq/g.
Biological Performance]
In vitro cytotoxicity test: cytotoxicity is not greater than grade 1;
Delayed hypersensitivity test: no delayed hypersensitivity;
Intracutmal reaction test: the difference between the average score of the test sample and the solvent control is 0;
Acute systemic toxicity test: no acute systemic toxicity;
Sub-chronic systemic toxicity test: no sub-chronic systemic toxicity
Amestest: negative for mutation
Mouse lymphoma gene mutation test: mutagenesis was negative.
[Applicable equipment]]This product is suitable for dental cutting or die casting equipment.
Contraindications]None
Storage Conditions]were stored at room temperature.
[Validity]]10 years.
[Note]]
1.Is made by professional technicians and used by professional doctors, and the products are in contact with oral mucosa after processing;
2. Do not inhale ceramic dust during the trimming process, and timely discharge exhaust gas and wear protective measures such as masks.
Contact: Andrew
Phone:
Tel:
E-mail: geaemed@gmail.com